Radiography: x-ray machine circuit - high voltage in Horses (Equis) | Vetlexicon
equis - Images

Radiography: x-ray machine circuit – high voltage

ISSN 2398-2977

Contributor(s) :



Radiography: x-ray machine circuit - high voltage

= annotation - select to reveal
  1. The high voltage circuit of the x-ray machine produces the large potential difference between the cathode (C) and anode (A) to accelerate.
  2. Thte auto transformer compensates for fluctuations in the national grid and allows a predetermined voltage to be delivered to the primary coil of the transformer.
  3. The step-up transformer increases the incoming voltage to the required kV.